The summary of the selected article appears at the bottom of the page. In order to get back to the contents of the issue this article belongs to you have to access the link from the title. In order to see all the articles of the archive which have as author/co-author one of the authors mentioned below, you have to access the link from the author's name.

 
       
         
    STUDIA PHYSICA - Issue no. SPECIAL%20ISSUE / 2001  
         
  Article:   IMPROVEMENT OF ISOTOPE RATIO MASS SPECTROMETRY ANALYSIS BY CHANGING THE AMPLIFIER CHANNEL AND BY AUTOMATIC DATA ACQUISITION.

Authors:  N. LUPÅžA, V. MIREL, STELA CUNA, ALINA N. NICA, ALEXANDRU CHIÅž.
 
       
         
  Abstract:  The goal of our work was to modernize the ATLAS M - 86 mass spectrometer in order to perform much accurate and precise isotopic ratio analyses. The improvement was realized by changing the old electronic units and by coupling the mass spectrometer to a personal computer (PC) - with the purpose to allow all of the spectrometer components to be controlled by the computer. Our research was focused on the ionic current measurement. The electronic components realizing this are the following: • the spectrometer interface • the control block of the spectrometer • the ionic currents measurement block  
         
     
         
         