The summary of the selected article appears at the bottom of the page. In order to get back to the contents of the issue this article belongs to you have to access the link from the title. In order to see all the articles of the archive which have as author/co-author one of the authors mentioned below, you have to access the link from the author's name.

 
       
         
    STUDIA CHEMIA - Issue no. 1 / 2020  
         
  Article:   SURFACE CHARACTERIZATION AND CYTOTOXICITY ANALYSIS OF THE TITANIUM ALLOYS FOR DENTISTRY.

Authors:  PAULA ARGENTINA JIMAN, MARIOARA MOLDOVAN, CODRUTA SAROSI, ALEXANDRINA MUNTEAN, ANDREEA SIMONA POP, VIORICA TARMURE, CATALIN POPA, AUREL GEORGE MOHAN.
 
       
         
  Abstract:  
DOI: 10.24193/subbchem.2020.1.12

Published Online: 2020-03-30
Published Print: 2020-03-30
pp. 149-162
VIEW PDF: FULL PDF

Three methods of treating the Ti and Ti-6Al-4V alloy surface used as dental implants have been tried both to increase the corrosion resistance and to create diffusion barriers, which prevent the release of toxic ions in the body, at the separation surface, implant living tissue. The anticorrosive performances of the experimental oxide films, realized on the pure Ti and Ti-6Al-4V alloy were evaluated electrochemically by cyclic voltammetry after immersion of samples in Ringer’s solution at 37°C. The A2-Ti sample (Ti-6Al-4V alloy anodized with H3PO4, 1N, and citric acid, 20 g/l electrolyte solution) shows lower viability values, compared to the other samples, with a tendency to decrease after 48h of incubation. The cytotoxic effect is slightly higher in A1-Ti (Ti-6Al-4V alloy anodized with 0.6% H3PO4 electrolyte solution) compared to Ti over a period. This can be attributed to the presence of aluminium and vanadium. The results revealed that the specific processing of titanium and titanium alloy (Ti-6Al-4V) by obtaining an oxide layer influence the toxicity that is stabilize and decrease with time, which makes to be used in dental implantology.

Keywords: Ti-6Al-4V surface treatment, electrochemical processing, dental implant, cytotoxicity, SEM
 
         
     
         
         